Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa 2433-11: (1 year) CAL VER /5500 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 7 NUMBER OF LINES: 80 CONFIGURATION: Fluke 5500A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # YOKOGAWA Bulletin 2433-E # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # NONE # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 STD Fluke 5500A/COIL 1.003 ASK- R P F V 1.004 ASK+ K X 1.005 HEAD PRELIMINARY INSTRUCTIONS 1.006 DISP Warm-up time: NONE 1.006 DISP Temperature : 18 - 28 degC 1.006 DISP Humidity: <80% 1.007 DISP [32] WARNING 1.007 DISP HIGH VOLTAGE is used or exposed during the performance 1.007 DISP of this calibration. DEATH ON CONTACT may result if 1.007 DISP personnel fail to observe safety precautions. 1.008 HEAD {AC VOLTAGE} 1.009 DISP Connect 5500A to the UUT as follows: 1.009 DISP [32] 5500A NORMAL HI to UUT P2 1.009 DISP [32] 5500A NORMAL LO to UUT P1 1.010 DISP Set the rotary switch to Vrms. 1.011 HEAD AC VOLTAGE: {200V Range} 1.012 5500 200 200.0V 1% 0.5/ 60H SI 2W 2.001 HEAD AC VOLTAGE: {600V Range} 2.002 5500 600 600.0V 1% 0.5/ 60H SI 2W 3.001 HEAD {AC CURRENT} 3.002 DISP Connect the 5500A/COIL to the 5500A AUX HI and LO 3.002 DISP terminals and clamp the UUT around the coil. 3.003 DISP Set the rotary switch to Arms. 3.004 HEAD AC CURRENT: {20A Range} 3.005 5500 20 20.00A 1% 0.5/ 60H SI T5 4.001 HEAD AC CURRENT: {200A Range} 4.002 5500 200 200.00A 1% 0.5/ 60H SI T5 5.001 HEAD {AC POWER} 5.002 DISP Set the rotary switch to kW. 5.003 HEAD AC POWER: {20kW Range} 5.004 M550 100V SI 5.005 5500 100A 60H SI S T5 5.006 ACC 10.00kW 0.0938U 5.007 MEMI Enter the UUT Power reading (in kilowatts). 5.008 MEME 5.009 MEMC 20 kW 2% 1/ 6.001 HEAD AC POWER: {200kW Range} 6.002 M550 500V SI 6.003 5500 200A 60H SI S T5 6.004 ACC 100.0kW 0.3U 6.005 MEMI Enter the UUT Power reading (in kilowatts). 6.006 MEME 6.007 MEMC 200 kW 2% 1/ 7.001 END